- Description
- Specifications
- Documents
Siglent 5615-12-P Secondary Reference Temperature Probe
Features and Specifications
- Temperature range: 196 to 420 Celcius
- Sheath: 305 mm (12 in) in length and 6.35 mm (0.250 in) in diameter
- Sheath Material: Inconel 600
- Maximum immersion length: 254 mm (10 in)
- Lead-wire cable type: 4-wire PTFE insulated with PTFE jacket, 22 AWG
- Lead-wire length: 183 cm (72 in or 6 ft)
- Termination: "P" (INFO-CON for 1523 or 1524)
- Reference-grade platinum sensing element
Compatible With
- 1523 or 1524 Handheld Thermometer Readout
- 1529 Chub-E4 Thermometer
- 1502A Tweener PRT Readout
5615-12-P Overview
The 5615-12-P (previously Hart Scientific) is a Platinum Resistance Thermometer (PRT) featuring an Inconel 600 sheath that measures 305 mm (12 in) in length and 6.35 mm (0.250 in) in diameter. It serves as a secondary reference temperature standard, bridging the gap between top-tier laboratory standards and industrial or secondary lab settings, with an accuracy of +/- 0.012 Celsius at 0.01 Celsius. The -P denotes the termination of the lead cable as INFO-CON.
The thermometer's element is made from reference-grade platinum wire (99.999% pure) for superior stability. The wire is coiled and placed in a mandrel, providing uniform support that nearly eliminates hysteresis. Its four-wire current-potential electrical configuration removes the effects of lead-wire resistance.
This Inconel-sheathed probe features a fully supported sensing element, enhancing durability compared to SPRTs. The element is housed in an ultrahigh-purity ceramic case with a hermetic glass seal, which improves output stability by preventing moisture and contaminants from entering.
INFO-CON Connectors Ensure Correct Temperature Conversion
Inside the INFO-CON, a memory chip keeps calibration information for the attached probe. Simply plugging in the probe uploads the information to the readout, ensuring the correct temperature conversion for accurate, hassle-free measurements.
5615 Series
The 5615 Series probes come calibrated with ITS-90 coefficients, making it compatible with many excellent readout devices, including Fluke Calibration’s 1529 Chub-E4, 1560 Black Stack, and 1502A Tweener. It bridges the gap between a 100-ohm industrial RTD and an SPRT.
Specifications for the 5626-00-X models are as follows, where 6,9,12 = Sheath and X = Termination:
Sheath Length:
- 5615-6-x: 152 mm x 4.76 mm (6.0 in x 0.188 in)
- 5615-9-x: 229 mm x 4.76 mm (9.0 in x 0.188 in)
- 5615-12-x: 305 mm x 6.35 mm (12.0 in x 0.250 in)
Termination:
- A (INFO-CON for 914X)
- B (bare wire)
- D (5-pin DIN for Tweener Thermometers)
- G (gold pins)
- J (banana plugs)
- L (mini spade lugs)
- M (mini banana plugs)
- P (INFO-CON for 1523 or 1524)
- S (spade lugs)